How to use the HemaClear® Pressure Table:

1: Download the pdf.

2: Note which model is used. (example: HemaClear®/40) Choose the Table Section that matches the color the surgeon is using (example: HemaClear®/40 - YELLOW). Measure (have surgeonmeasure) the distance from the tips of the fingers/toes to the final location where the HemaClear® ring is placed. Find the column in the Table that corresponds to this distance. Measurethe limb circumference at the site of the HemaClear¬Æ ring final position. Find the row in the Table that corresponds to the limb circumference at the final positionof the HemaClear®. The number in the column-row intersect is the pressure in mm Hg that the HemaClear® ring applies to the skin surface.NPD ‚Äì Non Physiological Dimension.